WWW > IMOP > Regional Instrument Centresl
 
RIC at Beijing, China (Region II)
(Information on this page is based on a report from the RIC received in August 2013. Please refer to the RIC's webpage for the latest information.)
 
General Information
Address: NO.46 ZHONGGUANCUN NANDAJIE,HAIDIAN DISTRICT,BEIJING,CHINA
Website: http://www.wmic.cn/
Contact person: LING BING
Title:  Ms
Email: linbing  @ cma.gov.cn
Tel:+0086 10 68400924
Fax:+0086 10 68400924
Calibration capabilities: Temperature, Pressure, Humidity, Wind, Precipitation, Solar Radiation
 
Specific Information on instrument calibration capabilities
Temperature measurement (thermometer)
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body
Liquid thermometer,
Digital thermometer

-60°C-+80°C

Platinium Resistance Thermometer
0.007 °C
2 Nov. 2012
National Institute of Metrology,China
Status of accreditation: unknown
 
Pressure measurement (barometer)
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body
Barometer,Pressure sensor,Digital Barometer,Piston gage

60hPa-1750hPa

Taylor Hobson roundness instrument,
Mitutoyo Legex707 Cmm

1 Pa at 1000 hPa
14 Jan. 2008
Institute for National Measurement Standards, Canada
Status of accreditation: unknown
 
Humidity measurement (hygrometer)
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body
Humidity sensor

DEW:-70°C-+50°C

Humidity Generator
0.15°C
17 June 2013
National Institute of Metrology,China
Status of accreditation: unknown
 
Wind measurement (anemometer)
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body
Pitot tube,
Anemometer,
Wind vane

0.2m/s-70m/s

Pitot tube
0.1%
23 Jul. 1987
NPL
Status of accreditation: unknown
 
Precipitation (rain gauge)
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body
Rain gauge

0mm-10mm

Standard Capacity Measures(glass)
0.1%

30 Jan. 2011

National Institute of Metrology, China
Status of accreditation: unknown
 
Solar Radiation (Pyrheliometer, etc.)
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body
Absolute pyrheliometer,
Pyrheliometer,
Pyranometer,
Net pyrradiometer

0W/m2-1400W/m2

Absolute pyrheliometer
0.25%
14 Oct. 2010
World Radiation Centre (Davos, Switzerland)
Status of accreditation: unknown

 

* A CMC (calibration and measurement capability) is the smallest uncertainty of measurement that can be expected to be achieved by the RIC during a calibration. This CMC is evaluated by the RIC itself and described in the scope of accreditation of the RIC, if available.

 