General Information
Address: Slovak Hydrometeorological Institute, Jeséniova 17, 833 15 Bratislava, Slovakia
Website: http://www.shmu.sk/en/?page=1541
Contact person: Lenka Leštinská
Email: lenka.lestinska@shmu.sk
Tel: 00421 2 59 415 124
Fax:
Calibration capabilities: Temperature, Relative Humidity, Atmospheric Pressure, Wind, Precipitation, Air Quality
 
Specific Information on instrument calibration capabilities
Temperature
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body

In-Glass thermometers

-30 °C to +40 °C

In-glass thermometer

0.2 °C

2017

SMU

Platinum resistance thermometers

-30 °C to +40 °C

SPRT + resistance bridge

0.05 °C

2017

SMU

Mechanical thermometers

-30 °C to +40 °C

PRT + data collector

1 °C

2017

SHMU

Electronic thermometers

-30 °C to +40 °C

PRT + data collector

0.3 °C

2017

SHMU

Status of accreditation (date of the latest accreditation): 2013
Link to the Certificate of Accreditation
Accreditation body:  Slovak National Accreditation Service (SNAS)
 
Relative Humidity
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body

RH sensors

10 to 97 %

Dew point standard

2.5 %

2017

CMI

Hygrographs

20 to 95 %

Capacitance RH sensor in climatic chamber

3 %

2017

SHMU

Mechanical hygrometers

20 to 95 %

Capacitance RH sensor in climatic chamber

5 %

2017

SHMU

Status of accreditation (date of the latest accreditation): 2013
Link to the Certificate of Accreditation
Accreditation body: Slovak National Accreditation Service (SNAS)
 
Atmospheric Pressure
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body

Absolute pressure

650 to 1050 hPa

Calibrator DHI PPC2+

0.1 hPa

2017

SMU

Status of accreditation (date of the latest accreditation): 2013
Link to the Certificate of Accreditation
Accreditation body: Slovak National Accreditation Service (SNAS)
 
Wind
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body

Anemometer

0.2 to 40 m/s

Thermal and rotating anemometer,

Pitot tube

0.1 m/s (BMC)

2017

CMI

Status of accreditation (date of latest accreditation): 2013
Link to the Certificate of Accreditation
Accreditation body: Slovak National Accreditation Service (SNAS)
 
Precipitation
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body

Tipping bucket

 

Volumetric flask

10 ml at 500 ml

2005

SMU

Weight rain gauge

1g to 7 kg

Set of weights

1 g

2017

SLM

Status of accreditation (date of latest accreditation): 2013
Link to the Certificate of Accreditation
Accreditation body: Slovak National Accreditation Service (SNAS)
 
Air Quality
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body

NOx analyzers

(0.5 to 1000) nmol/mol

Standard analyzer

(0.04 hm + 1.3) nmol/mol

2016

CHMI

SO2 analyzers

(0.5 to 1000) nmol/mol

Standard analyzer

(0.05 hm + 1.3) nmol/mol

2016

CHMI

O3 analyzers

(0.5 to 500) nmol/mol

Standard analyzer

(0.03 hm + 1.2) nmol/mol

2016

CHMI

CO analyzers

(0.04 to 50) μmol/mol

Standard analyzer

(0.02 hm + 0.4) μmol/mol

2016

CHMI

Status of accreditation (date of latest accreditation): 2013
Link to the Certificate of Accreditation
Accreditation body: Slovak National Accreditation Service (SNAS)
 

* A CMC (calibration and measurement capability) is the smallest uncertainty (k=2) of measurement that can be expected to be achieved by the RIC during a calibration under normal conditions. This CMC is evaluated by the RIC itself and described in the scope of accreditation of the RIC, if available.

 
Further Information
 
 



top

Top






©2017 World Meteorological Organization
, 7bis, avenue de la Paix, Case postale No. 2300, CH-1211 Geneva 2, Switzerland 
Tel.: +41(0)22 730 81 11 - Fax: +41(0)22 730 81 81
Disclaimer | UN System | Accessibility

 

  Extranet Homepage Website Public Comms