WWW > IMOP > Regional Instrument Centresl
 

RIC of Ljubljana, Slovenia (Region VI)

(Information on this page is based on a report from the RIC received in August 2013.
Please refer to the RIC's webpage for the latest information.)
 
General Information
Address: Vojkova 1b, 1000 Ljubljana, Slovenia
Website: N/A
Contact person: GROSELJ Drago
Title: Head of Calibration laboratory
Email:drago.groselj @ gov.si
Tel: +386 1 478 4100
Fax:+386 1 478 4050
Calibration capabilities: Temperature, Pressure, Humidity
 
Specific Information on instrument calibration capabilities
Temperature measurement (thermometer)
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body
Platinum resistance thermometer

0.01°C

Triple point of water cell, Isotech, 811L-32
0.7 mK
(Level of confidence approximately 95 %)
21 Nov. 2010
LMK (Slovenia)
29.7646°C
Gallium melting point cell, Isotech, ITL-M-17401
0.9 mK
(Level of confidence approximately 95 %)
21 Nov. 2010
-40~50 °C
Standard platinum resistance thermometer, Isotech SPRT 419
Liquid bath, Isotech 915 LW

13 mK
(Level of confidence approximately 95 %)
31 Aug. 2012
LMK (Slovenia)
Status of accreditation: date of latest accreditation - 12 April 2013
Link to the accreditation certificate
Accreditation body:  Slovenian Accreditation (SA)
 
Pressure measurement (barometer)
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body
Digital barometer

500 to 1100 hPa

        Digital piston gauge DH24610
Generator DH type PPC1

6 Pa + 1.3 10-5 P
        (Level of confidence approximately 95 %)

1 May 2013
CMI (Czech Republic)
Status of accreditation: date of latest accreditation - 12 April 2013
Link to the accreditation certificate
Accreditation body: Slovenian Accreditation (SA)
 
Humidity measurement (hygrometer)
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body

Electronic hygrometer

10~95 %

(in T range of -10°C to 50°C)

Chilled mirror hygrometer Mitchell S4000 Climatic
Humidity generator Thunder Scientific 2500

1.6% RH (40°C ≤ Ta ≤ 50°C)
1.8% RH (20°C ≤ Ta < 40°C)
2.2% RH (10°C ≤ Ta < 20°C)
2.8% RH (0°C ≤ Ta < 10°C)
3.4% RH (-10°C ≤ Ta < 0°C)
(Level of confidence approximately 95 %)

29 August 2012
LMK (Slovenia)
Status of accreditation: date of latest accreditation - 12 April 2013
Link to the accreditation certificate
Accreditation body: Slovenian Accreditation (SA)

 

* A CMC (calibration and measurement capability) is the smallest uncertainty of measurement that can be expected to be achieved by the RIC during a calibration. This CMC is evaluated by the RIC itself and described in the scope of accreditation of the RIC, if available.

 